Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8: CAL ADJ /5100 INSTRUMENT: Fluke 10: CAL ADJ /5100 INSTRUMENT: Fluke 11: CAL ADJ /5100 INSTRUMENT: Fluke 12: CAL ADJ /5100 INSTRUMENT: Fluke 12B: CAL ADJ /5100 INSTRUMENT: Fluke 18: CAL ADJ /5100 DATE: 19-Aug-96 AUTHOR: Fluke Corporation REVISION: $Revision: 2.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 64 CONFIGURATION: Fluke 5100B ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD CALIBRATION ACCESS PROCEDURE 1.002 DISP Remove any input signals to the meter and turn off. 1.002 DISP 1.002 DISP Remove the four screws on the back and separate the 1.002 DISP case bottom and case top. 1.002 DISP 1.002 DISP Notice that when you do so the battery remains in the 1.002 DISP case bottom and power to the meter is disconnected. 1.003 DISP Observing correct polarity, connect a 9V battery to the 1.003 DISP battery contacts using easy hook junpers or alligator 1.003 DISP clip leads. 1.003 DISP 1.003 DISP [32] NOTE 1.003 DISP To avoid stretching or bending the battery contacts, 1.003 DISP connect leads to the base of the contacts. 1.004 HEAD EQUIPMENT SETUP 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Allow the UUT to stabilize. 1.006 HEAD CALIBRATION PROCEDURE 1.007 DISP Move the Slide-Switch to the center position. 1.007 DISP Press RANGE to select the 4V Range. 1.008 5100 4V S 2W 1.009 DISP Adjust R4 reading of 4.000+/-0.003V. 1.010 END